C.S. Certification Exam, SPD Test, Sterile
Processing Study Material For Certification Exam
Which area is the first to accept responsibility for all supplies in the facility?
Correct Ans - Receiving
Which type of inventory distribution system requires the use of a pre-
determined inventory list? Correct Ans - Par Level
Just-in-time distribution systems are also known as Correct Ans -
Stockless Distribution
Ordering the optimum quantity of an item based on historical data is known
as Correct Ans - Economical Order Quantity
Which of the following are elements of a distribution system? Correct Ans
- patient charging, storage, and receiving
The most efficient distribution system is the Correct Ans - Total
exchange cart
The computerized locked supply system is a variation of the Correct Ans -
Par Level
Which distribution system requires the using department to be responsible
for maintaing inventory levels? Correct Ans - Demand
An ongoing inventory management system is known as a Correct Ans -
Perpetual Inventory System
In which distribution system are requisition filled and delivered as they are
received? Correct Ans - Demand
Inventory performance" (a rough indication of the effectiveness of the
inventory system) is evaluated by Correct Ans - percent of back order,
availability of critical items, and inventory turnover
,Which of the following is the MOST efficient and cost-effective method for
inventory control in the SPD ordering and data generation? Correct Ans -
Computerized Inventory Control System
The temperature in the sterile storage are should not exceed Correct Ans
- 75º F (24º C)
Shelving in the sterile storage area must be kept how far from the outside
walls? Correct Ans - 2 inches
The thickness of a sterility maintenance cover should be Correct Ans -
2 to 3 mils
The minimum number of air exchanges per hour in the sterile storage area
should be Correct Ans - 4
The relative humidity in the sterile storage area should not exceed Correct
Ans - 70%
Approximately how much space should be left between and item and the
inside edges of a peel pouch? Correct Ans - 1 inch
What is the maximum size and weight of a textile pack? Correct Ans -
12" x 12 x 20" 12 pounds
Which of the following packaging materials is acceptable for low-temperature
gas plasma and ozone sterilization Correct Ans - Tyvek
Basin sets should be prepared for sterilization with Correct Ans - an
absorbent material between each basin
Tamper-evident seal on rigid containers include Correct Ans - latches
and locks
What is the acceptable temperature range for the prep and packaging area?
Correct Ans - 68º F-73º F (20º-23º C)
The preparation and packaging area must have an air exchange rate of
Correct Ans - 10 per hr
,If a non-perforated tray is used for the sterilization of instrument sets, it must
be Correct Ans - Tilted on the sterilizer rack
The ideal humidity level in the preparation and packing area should be
Correct Ans - 50%
For sterilization, rigid containers must be placed Correct Ans - flat
Which of the following packaging materials is acceptable for steam
sterilization? Correct Ans - Non-woven wraps
Re-passivation of an instrument can be achieved by Correct Ans -
routine lubrication
What is the most difficult part of a hemostat to clean? Correct Ans -
box lock
Which of the following lubricants is recommended for use on surgical
instruments? Correct Ans - petroleum-based
What type of finish on a surgical instrument is used for laser procedures?
Correct Ans - ebonized
What is the final stage in manufacturing a surgical grade instrument?
Correct Ans - passivation
Which series of stainless steel is used is to manufacture malleable retractors?
Correct Ans - 300
The weakest part of the instrument is/are the Correct Ans - box locks
The layer on a surgical instrument created by the passivation process is called
Correct Ans - chromium oxide
A crack in a hand-held instrument would most likely be found in the
Correct Ans - box lock
, The appropriate sterilization cycle for an autoclave load containing a basin
set, a minor surgery set, and twelve towel packs is Correct Ans - 270 F
(132 C ) for 4 minutes, 30 psi and 20 minutes drying time
Cotton gauze cannot be used in which sterilization process? Correct Ans -
low temperature gas plasma
Which agencies are involved with the regulation and control of ethylene oxide
when it's used as a sterilant? Correct Ans - environmental protection
agency, food and drug administration, and centers for disease control and
prevention
Which of the following sterilization methods would be appropriate for
petrolatum gauze dressings? Correct Ans - dry heat
To eliminate the bioburden on metal surgical instrument within the SPD, the
MOST cost-efficient sterilization method is generally considered to be
Correct Ans - steam under pressure
There are two pressure gauges on a steam sterilizer. These gauges measure
the pressure in the Correct Ans - jacket and chamber
What is the maximum number of 100% EO cartridges permitted by OSHA to
be stored outside of a flammable goods cabinet? Correct Ans - 10
At the end of the sterilization cycle, the EO cartridges and the gloves used to
handle them should be Correct Ans - Aerated
Item to be sterilized in low-temperature gas plasma must be packaged in:
Correct Ans - Non-cellulose-based material
Generally, the EO biological test pack is placed in the Correct Ans -
Geometric center of the load
Which of the following devices are compatible with ozone sterilization?
Correct Ans - stainless steel instruments
Routine BI testing for steam sterilizers should be performed at least
Correct Ans - weekly
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card or Stuvia-credit for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller Studyhall. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for $15.99. You're not tied to anything after your purchase.