Answers Guaranteed Success 2024/2025
Verified that the sterilization conditions within the load were adequate to kill a
known number of bacterial spores. - correct answer Biological monitoring
Biological monitoring - correct answer - is the only way to measure the lethality of
a sterilization process.
Geobacillus stearothermophilus - correct answer The spore that is most resistant
to steam sterilization is
Indicator tape - correct answer Type 1 process indicators
Bowie-dick - correct answer Type 2: indicators for specific tests
Single-variable indicators - correct answer Type 3
Type 4 indicators - correct answer Multi-variable indicators
Integrating indicators - correct answer Type 5
Emulating indicators - correct answer Type 6
, Biological monitoring - correct answer Verifies that the sterilization conditions
within the load were adequate to kill a known number of bacterial spores.
60-65 degrees/ 30-60% - correct answer Decon temperature and humidity:
68-73 degrees/ideally 50% no less than 35% - correct answer Prep & pack
temperature and humidity:
68-73 degrees/ 30%-60% - correct answer Sterilization temperature and humidity:
68-75 degrees/35%-70% - correct answer Sterile Storage temperature and
humidity:
60-65 degrees/ 30-60% - correct answer HLD area temperature and humidity:
4 per hour - correct answer Air exchanges for sterile storage?
HLD - correct answer According to Spauldings classification of medical devices, a
semi critical item requires which form of disinfection?
Light-weight items - correct answer Paper-plastic pouches can be used for
Dwell time - correct answer Critical parameters for heat sealing paper-plastic
pouches include temperature, pressure, and