Explain how to calculate air changes per hour. - Answer✔✔-Room dimension: length x width x height =
cubic feet.
Total CFM from all HEPA filters in room.
Total CFM / Room volume x 60 = ACPH.
Explain how to calculate air changes per hour from PEC - Answer✔✔-Length x width x fpm = CFM.
Total CFM divided by room volume x 60 = ACPH per that particle PEC.
Explain how to determine minimum CFM required from ceiling HEPA filters in HD room. - Answer✔✔-
Take minimum 30 ACPH (or whatever ACPH for that ISO Class) / 60 = 0.5 and multiply by the room
volume (500 cu. ft.) =250 CFM total HEPA filtered air supply.
Explain how to calculate the quality of particle counts in a room. - Answer✔✔-Take room dimensions of
300" length x 413" width to achieve 123,900 in2 and divide by 144 = 860.417 ft2.