NOMAD Pro's smaller diameter cone, combined with the use of DC power, and
primary shielding greatly reduce a patient X-ray exposure.
True
False - ANSWERTrue
The NOMAD may be operated solo by a technician while they are still in training on
how to use the device.
True
False - ANSWERFalse
Because NOMAD Pro shielding limits the exposure, it is permissible to have other
people in the exam room during exposures without any precautions.
True
False - ANSWERFalse
The NOMAD Pro should be routinely cleaned using disinfectant wipes, or some other
non- acetone cleaner to maintain cleanliness.
True
False - ANSWERTrue
The backscatter shield is adjustable to accommodate positioning devices and other
special needs but doing so requires additional precautions.
True
False - ANSWERTrue
If the NOMAD Pro has been dropped, it should be returned to Aribex for a safety
check.
False
True - ANSWERTrue
The NOMAD Pro should normally be held with two hands for stability whenever
possible.
True
False - ANSWERTrue
Do not enable/ready the NOMAD Pro until the patient and film / sensor are
positioned and ready to avoid unintended firing of the NOMAD Pro.
False
True - ANSWERTrue
When there is not enough battery charge left to take an exposure, the charge
indicator bars will fade away and the recharge battery alarm message appears on
the screen and the NOMAD Pro will not allow an exposure to begin.
, True
False - ANSWERTrue
An exposure can be prematurely terminated for any reason by releasing the trigger
at any time during the exposure.
True
False - ANSWERTrue
The Handset can be left on the Charging Cradle overnight without harming the
batteries.
True
False - ANSWERTrue
The Handset should be detached from the NOMAD Pro at the end of each day's use.
True
False - ANSWERTrue
The Handset should be detached from the NOMAD Pro at the end of each day's use.
True
False - ANSWERTrue
When the NOMAD Pro is powered on and in the settings screen, a single quick pull
of the trigger will enable the device, the green LED will illuminate and the display will
indicate the device is ENABLING and then the _______ message will appear on the
screen (meaning the next trigger pull will fire X-rays).
Not ready
Ready - ANSWERReady
When the NOMAD Pro is not in service, it is advisable to ensure that the device is
securely stored and the X-rays are __________ so that unauthorized / unintended
use or damage does not occur.
Locked
Unlocked - ANSWERLocked
An alarm indicating a partial exposure (when the trigger was not depressed through
the entire timer setting - either on purpose or accidentally) is indicated by a series of
beeps and the _________ message on the screen until a button is pressed.
Complete
Incomplete - ANSWERIncomplete
If left unused NOMAD Pro will automatically shut down after approximately
________ minutes. Or the NOMAD Pro can be turned by pressing the POWER
button.
1
2-3