Which division would have primary jurisdiction over a vascular graft with an antibiotic based on
primary mode of action?
A. CDER
B. CBER
C. CDRH
D. OCP - ** VERIFIED ANSWERS **✔✔C
A company wants to modify its legally marketed device such that the modification does not affect the
intended use or alter the fundamental scientific technology of the device. If the design outputs of the
modified device meet the design input requirements, this change would be best filed as a(n):
A. Special 510(k)
B. Abbreviated 510(k)
C. Traditional 510(k)
D. De novo 510(k) - ** VERIFIED ANSWERS **✔✔A
Under the statutory violations, failure to meet 510(k) requirements for a device that is required to
have a 510(k) and is in commercial distribution is considered to be:
A. Adulteration.
B. Improper use
C. Misbranded
D. Fraudulent - ** VERIFIED ANSWERS **✔✔C
A company's competitor is marketing a Class II suture which dissolves during the third week of use.
The company's current product has to be removed by a physician. However, a change in weaving
configuration gives this product the same dissolving time as the competitor's. When can the
company's new suture be marketed?
,A. This requires a new 510(k) since significant change in product instructions might
affect efficacy.
B. After submission in a periodic report
C. After reporting clinical studies in an annual report
D. After submission of labeling change - ** VERIFIED ANSWERS **✔✔A
Which of the following is exempt from GMP/QSR regulations?
A. Remanufacturers
B. Custom device manufacturers
C. Repackagers
D. Component manufacturers - ** VERIFIED ANSWERS **✔✔D
A physician reports to a manufacturer that a patient was hospitalized with acute sepsis after
treatment with an approved device. This side effect is not listed in the package insert. This event
must be reported by the manufacturer to FDA no later than:
A. 5 calendar days
B. 15 calendar days
C. 30 calendar days
D. The next quarterly or annual report - ** VERIFIED ANSWERS **✔✔C
If a device failure is occurring with greater than expected frequency and investigation of the problem
implicates improper use by the end user, which of the following typically occurs?
A. The labeling is revised.
B. The product is recalled.
C. The product is redesigned.
D. A "Dear Doctor" letter is issued. - ** VERIFIED ANSWERS **✔✔A
A handling and storage system for medical devices must always include:
A. Procedures for rotation of stock
, B. Separate rooms or cages for release and quarantine products
C. Procedures for receipt and transfer of product
D. Environmentally controlled areas for products with shelf life - ** VERIFIED ANSWERS **✔✔C
You have modified your 510(k) cleared device with a special 510(k). In which of the following cases
would you need to create a new listing for the device?
A. You have added new sizes and shapes in the product portfolio.
B. You have changed the material composition of the device.
C. You have changed the package of the device.
D. None of the above. - ** VERIFIED ANSWERS **✔✔D
According to the QSR, when an investigation of a complaint is conducted all of the following are
requirements for inclusion in the record of the investigation EXCEPT:
A. The dates and results of the investigation
B. The nature and details of the complaint
C. Changes in procedures correcting quality problems
D. Any reply to the complainant - ** VERIFIED ANSWERS **✔✔C
The QSR calls for the manufacturer of finished devices to carry out all of the following EXCEPT:
A. Conduct quality audits by individuals who do not have direct responsibility for the operation
being audited
B. Audit operations annually
C. Document the dates and results of quality audits and re-audits
D. Have findings reviewed by management responsible for the matters audited - ** VERIFIED
ANSWERS **✔✔B
Which of the following subsystems is NOT required by FDA in order to implement and maintain a
Quality System?
A. Production and process controls
B. Test and control article characterization
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card or Stuvia-credit for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller Fordenken. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for $16.99. You're not tied to anything after your purchase.