What do we need to build an EM?
1. Vacuum
2. Emission source
3. Lenses
4. Apertures
5. Detectors
1. Vacuum
A TEM is kept permantaly under vacuum, unless it’s being repaired or serviced. There is
always a vacuum because:
electrons interacted with the atoms in air which causes scattering which causes a short
path length in air (20 cm in air and 2000 km in vacuum, 300 kV).
Next to that the vacuum keeps the specimen clean.
Access to the inside of the column via an airlock system, which can be pumped separately.
The airlock prevents degrading the vacuum of the column when inserting a sample. The
sample is initially pumped from atmospheric pressure to an intermediate vacuum, and then
that intermediate vacuum is mixed with the high column vacuum when the sample is fully
inserted. This reduces the amount that the column must be pumped to maintain high vacuum.
Different parts of the microscope are under different vacuum. The column and electron source
are under a very ultra-high vacuum.
, There are different types of pumps to creates different vacuums.
Exhaust pumps
Trapping pumps
With a IGP you remove air by emitting electrons from a cathode and then by ionizing air
molecules. The ion is going to move to the grounded cathode plate with a high velocity.
When it hits that plate, the ions will become buried within the cathode or they will sputter
cathode material onto the walls of the pump. This sputtered material then removes the gas
molecules by absorbing. You can only use a IGP when you already have a high vacuum.
2. Electron sources
The electron gun needs to
1. extract electrons from a tip
2. accelerate the electrons
There are different ways for extracting electrons from a material:
Thermionic sources produces electrons when heated
Gives the electrons sufficient energy to overcome the natural barrier that prevents
them from leaking out from the surface. The barrier is called the work function .
You can only use materials with:
high melting point: Tungsten (W) filament
cheap butbreaks very fast
low work function : lanthanum hexaboride
(LaB6) crystals.
thermionic sources are less monochromatic (energy
different, different)
Les avantages d'acheter des résumés chez Stuvia:
Qualité garantie par les avis des clients
Les clients de Stuvia ont évalués plus de 700 000 résumés. C'est comme ça que vous savez que vous achetez les meilleurs documents.
L’achat facile et rapide
Vous pouvez payer rapidement avec iDeal, carte de crédit ou Stuvia-crédit pour les résumés. Il n'y a pas d'adhésion nécessaire.
Focus sur l’essentiel
Vos camarades écrivent eux-mêmes les notes d’étude, c’est pourquoi les documents sont toujours fiables et à jour. Cela garantit que vous arrivez rapidement au coeur du matériel.
Foire aux questions
Qu'est-ce que j'obtiens en achetant ce document ?
Vous obtenez un PDF, disponible immédiatement après votre achat. Le document acheté est accessible à tout moment, n'importe où et indéfiniment via votre profil.
Garantie de remboursement : comment ça marche ?
Notre garantie de satisfaction garantit que vous trouverez toujours un document d'étude qui vous convient. Vous remplissez un formulaire et notre équipe du service client s'occupe du reste.
Auprès de qui est-ce que j'achète ce résumé ?
Stuvia est une place de marché. Alors, vous n'achetez donc pas ce document chez nous, mais auprès du vendeur annickkooij. Stuvia facilite les paiements au vendeur.
Est-ce que j'aurai un abonnement?
Non, vous n'achetez ce résumé que pour 2,99 €. Vous n'êtes lié à rien après votre achat.