The Biotronik CRT feature, AV Delay & Negative Hysteresis, is designed to: (choose the best answer)
Provide a lower rate when the intrinsic rate is faster than the lower rate
This is not a Biotronik feature
Optimize A-V
Reduce RV pacing - AN...
The Biotronik CRT feature, AV Delay & Negative Hysteresis, is designed to: (choose the best answer)
Provide a lower rate when the intrinsic rate is faster than the lower rate
This is not a Biotronik feature
Optimize A-V
Reduce RV pacing - ANSWERSOptimize A-V
What additional steps are required when conducting a CRT follow-up (3-leads) compared to an ICD
follow-up (2-leads)? (select all that apply)
A. Evaluate Cardiac Compass/trend information
b. Assess CRT pacing: if less than 90% investigate more
c. Evaluate patients AV conduction to determine if MVP would be appropriate
d. Evaluate Wavelet template since Auto-Collection is inactive
e. Evaluate LV lead performance for impedance and threshold - ANSWERSB
D
E
, AdaptivCRT is a dynamic, physiologic pacing algorithm which enhances CRT by adjusting CRT parameters
automatically to changes in the patient's activity and conduction. It can be programmed to provide
pacing by 1 of 3 different options: (choose all that apply)
Adaptive BiV
CardioSync
Adaptive BiV & LV
Non-Adaptive CRT
QuickOpt - ANSWERSAdaptive BiV
Adaptive BiV & LV
Non-Adaptive CRT
Which of these LV pace polarity options would maximize device longevity without Phrenic Nerve
stimulation:(LV Pace Polarity; Threshold; Phrenic Nerve Stimulation (PNS); Impedance)
A. LV Tip to LV Ring; 1.5V @
B. 0.4ms; No PNS; 825 ohms
c.LV Ring to LV Tip; 2.0V @ 0.4ms; Yes PNS; 1000 ohms
d LV Tip to RV Coil; 1. - ANSWERSLV Tip to LV Ring; 1.5V @ 0.4ms; No PNS; 825 ohms
A night heart rate of >85 bpm is suggestive of:
HF decompensation is impending
Too much stress and needs an anti-anxiety medication
Poor physical conditioning and needs exercise
A "My Pillow" prescription is needed - ANSWERSHF decompensation is impending
The percentage of pacing for a heart failure patient with a CRTD device should be at least:
Voordelen van het kopen van samenvattingen bij Stuvia op een rij:
√ Verzekerd van kwaliteit door reviews
Stuvia-klanten hebben meer dan 700.000 samenvattingen beoordeeld. Zo weet je zeker dat je de beste documenten koopt!
Snel en makkelijk kopen
Je betaalt supersnel en eenmalig met iDeal, Bancontact of creditcard voor de samenvatting. Zonder lidmaatschap.
Focus op de essentie
Samenvattingen worden geschreven voor en door anderen. Daarom zijn de samenvattingen altijd betrouwbaar en actueel. Zo kom je snel tot de kern!
Veelgestelde vragen
Wat krijg ik als ik dit document koop?
Je krijgt een PDF, die direct beschikbaar is na je aankoop. Het gekochte document is altijd, overal en oneindig toegankelijk via je profiel.
Tevredenheidsgarantie: hoe werkt dat?
Onze tevredenheidsgarantie zorgt ervoor dat je altijd een studiedocument vindt dat goed bij je past. Je vult een formulier in en onze klantenservice regelt de rest.
Van wie koop ik deze samenvatting?
Stuvia is een marktplaats, je koop dit document dus niet van ons, maar van verkoper Bensuda. Stuvia faciliteert de betaling aan de verkoper.
Zit ik meteen vast aan een abonnement?
Nee, je koopt alleen deze samenvatting voor €8,06. Je zit daarna nergens aan vast.