Explain how to calculate air changes per hour. - Answer-Room dimension: length x width x height = 
cubic feet. 
Total CFM from all HEPA filters in room. 
Total CFM / Room volume x 60 = ACPH. 
Explain how to calculate air changes per hour from PEC - Answer-Length x width x fpm = CFM. 
Total CFM divided by room volume x 60 = ACPH per that particle PEC. 
Explain how to determine minimum CFM required from ceiling HEPA filters in HD room. - An...