CER Review: Questions And Answers
Leak Testing should be performed: ✔️-After point-of-use cleaning
-Before manual cleaning
-After every procedure
Early detection of endoscope damage can protect the patient and:
✔️Reduce repair costs
Inadequate work space in the endoscope work area can be a cause of:
✔️Endoscope damage
When should the high-level disinfectant solution in an automated equipment
reprocessor (AER) be tested for the minimum effective concentration (MEC):
✔️Before each use
All flexible endoscopes should have a ventilation adaptor or water-resistant
cap in place when undergoing ethylene oxide sterilization. T/F
✔️False
Biological indicator monitoring is required for each cycle run in:
✔️Ethylene Oxide
When using a rigid endoscope, a dark image may be caused by damaged:
✔️-Illumination fibers
-Coupler
-Camera
This endoscope is often used to treat kidney stones ✔️Ureteroscope
Biofilm and fungus can form underneath fingernails. T/F ✔️True
Endoscopes that have been high-level disinfected should be stored on open
wire shelving. T/F ✔️False. A dedicated cabinet should be used to
store flexible endoscopes .
Low humidity in the work area can: ✔️Cause absorbent materials to
become excessively dry.
, The process of rendering a clinically-used devise safe and ready for reuse is
the definition of: ✔️Reprocessing
When performing a leak test, the water-resistant cap should be removed: T/F
✔️False. Always ensure that the cap is fully "on" and "locked"
The four primary stains used for bacterial identification: ✔️Gram (+-),
Acid Fast, Endospore Stain, Capsule Stain
Decontamination Area controls: ✔️-Temp: 60-65F
-Humidity: 20-60%
-Airflow: 10 negative air flows
Preparation / Disinfection Area controls: ✔️-Temp: 60-73F
-Humidity: 20-60%
-Airflow: 10 positive air flows
Storage Area controls: ✔️-Temp: 60-75F
-Humidity: Above 20%
-Airflow: 4 positive air flows
Inadequate work space in the endoscope work area can be a cause of:
✔️Endoscope damage
The AER should be placed in the decontamination area to ensure the
endoscopes are clean before sending them to the preparation area. T/F
✔️False. They should be located in a prep/disinfection area to reduce risks of
cross decontamination .
Detergents used to clean endoscopes should be: ✔️Non foaming and
free rinsing.
Enzymes will become ineffective in temperatures that exceed:
✔️140F
The most important source of information for determining the proper
chemicals to use for cleaning an endoscope is the: ✔️Chemical
Manufacturer's IFU
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller Studyhall. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for £10.15. You're not tied to anything after your purchase.