100% satisfaction guarantee Immediately available after payment Both online and in PDF No strings attached
logo-home
CAMRT MRI Questions and Answers £15.48   Add to cart

Exam (elaborations)

CAMRT MRI Questions and Answers

 8 views  0 purchase
  • Module
  • CAMRT MRI
  • Institution
  • CAMRT MRI

CAMRT MRI Questions and Answers How far away must the operator stand from the xray tube if not covered with lead? - CORRECT ANSWER-3 meters 90% of man made radiation to the population is related to? - CORRECT ANSWERMedical Radiography and Dental For each peak xray voltage, what lead equivalent...

[Show more]

Preview 3 out of 17  pages

  • August 20, 2024
  • 17
  • 2024/2025
  • Exam (elaborations)
  • Questions & answers
  • CAMRT MRI
  • CAMRT MRI
avatar-seller
CAMRT MRI Questions and Answers
How far away must the operator stand from the xray tube if not covered with lead? -
CORRECT ANSWER-3 meters

90% of man made radiation to the population is related to? - CORRECT ANSWER-
Medical Radiography and Dental

For each peak xray voltage, what lead equivalent is required?

100kV or less (o-100)
Between 101kV-149kV
150kV or greater (150+) - CORRECT ANSWER-0-100kV = 0.25mm of lead

Greater than 100kV but less then 150kV = 0.35mm of lead

150kV or greater = 0.50mm of lead

What protective equipment is required during interventional procedures - CORRECT
ANSWER-Full wrap around protective gowns of 0.50mm pb in the front panels and
0.25mm pb in the back panels

Thyroid shields with an equivalent of 0.50mm of pb

Lead drapes/curtains mounted to II of 0.25mm of pb at 100kv

What lead equivalent must protective gloves/ gauntlets possess? - CORRECT
ANSWER-0.25mm of pb

What lead equivalent must ceiling mounted lead screens and moveable shields
provide? - CORRECT ANSWER-0.50mm of pb

What are the annual radiation dose limits for radiation workers? - CORRECT ANSWER-
Whole body= 20mSv
Lens of eye= 150mSv
Skin = 500mSv
Hands = 150mSv
All other organs = 500mSv

For pregnant women an effective dose of 4 mSv must be applied for the remainder of
the pregnancy from all sources

What are the annual radiation dose limits to members of the public? - CORRECT
ANSWER-Whole body = 1mSv

,Lens of the eye= 15mSv
Skin = 50mSv
Hands = 50mSv
All other organs = 50mSv

*these also apply for students*

What are primary protective barriers? Give examples - CORRECT ANSWER-Primary
barriers provide shielding from the direct xray beam. Such as, the walls and floor where
the tube can be directed

-The wall behind the wall bucky
-Floor under the table

What are secondary barriers and give examples - CORRECT ANSWER-Provide
shielding against both leakage and scattered radiation. Walls that the primary beam is
not directed towards and the ceiling are secondary barriers.

The control booth is also a secondary barrier, the primary beam should never be
directed towards it.

What should the focal spot to skin distance be for:

-Mobile equipment
-Stationary equipment
-Radioscopic equipment - CORRECT ANSWER-Must not be less than:

-Mobile: 30cm
-Stationary: 38cm
-Radioscopic: 20cm (designed for special procedures that would be impossible at 30cm
or 38cm)

What are the maximum air kerma rates for radioscopic equipment:

-with automatic intensity control
-without AIC - CORRECT ANSWER-Not equipped with AIC: 50mGy/min

With AIC: 100mGy/min

The air kerma rate must not exceed ___ from scattered radiation of radioscopic
equipment - CORRECT ANSWER-2 mGy/h

HVL of Aluminum for given tube voltages? (70kV-150kV) - CORRECT ANSWER-70:
2.5mm
80: 2.9mm
90: 3.2mm

, 100: 3.6mm
110: 3.9mm
120: 4.3mm
130: 4.7mm
140: 5.0mm
150: 5.4mm

How frequently should IP's and cassettes be cleaned and inspected? - CORRECT
ANSWER-Cleaned: monthly
Visual inspection: Weekly

What are daily QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER--Warm
up (all)
-Meter operation (all)
-Equipment condition (all)
-System movements (fluoro)
-Visual assessment of electronic display devices (all)

What are weekly QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER--
Visual inspection of cleanliness of system (all)
-Laser film printer operation (all)
-CT # accuracy
-CT noise
-CT uniformity

What are monthly QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER--
Cassette, screen, IP plate cleaning (CR)
-Retake analysis (CR/DR)
-Electronic display device performance (all)
-Laser film printer operation (all)
-CT slice thickness
-CT # calibration
-CT # linearity

What should the monthly retake analysis be below? - CORRECT ANSWER-5%

What are quarterly QC tests for CR/DR/Fluoro/CT - CORRECT ANSWER--Collimator
operation (CR/DR/Fluoro)
-Interlocks (all) (on doors to prevent exposure if door is open)
-Compression devices (Fluoro)
-Table angling/motion (Fluoro)
-Fluoro protection devices (lead drapes = .25mm of pb)
-CT patient support movement
-CT spatial resolution
-CT low contrast detectability

The benefits of buying summaries with Stuvia:

Guaranteed quality through customer reviews

Guaranteed quality through customer reviews

Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.

Quick and easy check-out

Quick and easy check-out

You can quickly pay through credit card for the summaries. There is no membership needed.

Focus on what matters

Focus on what matters

Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!

Frequently asked questions

What do I get when I buy this document?

You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.

Satisfaction guarantee: how does it work?

Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.

Who am I buying these notes from?

Stuvia is a marketplace, so you are not buying this document from us, but from seller TIFFACADEMICS. Stuvia facilitates payment to the seller.

Will I be stuck with a subscription?

No, you only buy these notes for £15.48. You're not tied to anything after your purchase.

Can Stuvia be trusted?

4.6 stars on Google & Trustpilot (+1000 reviews)

79373 documents were sold in the last 30 days

Founded in 2010, the go-to place to buy revision notes and other study material for 14 years now

Start selling
£15.48
  • (0)
  Add to cart