Alarm switches / sensors shall be installed so that they can be _ and accessible for service and testing. 99-50 5.1.9.1 (15) - Answer-Removable
All master, area, and local alarm systems shall include separate _ indicators for each condition monitored. 99-50 5.1.9.1 (1) - Answer-Visual
All med...
med gas study guide worksheet 7 exam questions and
alarm switches sensors shall be installed so tha
Geschreven voor
Gas Master
Gas Master
Verkoper
Volgen
lectknancy
Ontvangen beoordelingen
Voorbeeld van de inhoud
Med Gas Study Guide Worksheet 7 Exam
Questions and Answers
Alarm switches / sensors shall be installed so that they can be _ and accessible for
service and testing. 99-50 5.1.9.1 (15) - Answer-Removable
All master, area, and local alarm systems shall include separate _ indicators for each
condition monitored. 99-50 5.1.9.1 (1) - Answer-Visual
All medical gas alarm systems shall have an _ audible indication for each condition
monitored. 99-50 5.1.9.1 (3) - Answer-Cancelable
Anesthetizing locations where moderate sedation, deep sedation or general anesthesia
is administered, what alarm panels would monitor these locations? 99-52 5.1.9.44 (2) -
Answer-Area Alarm Panel
Area alarm panels for medical gas systems shall indicate when line pressure in the area
being monitored increase or decreases by _ percent. 99-52 5.1.9.4.2 - Answer-20%
Areas for anesthetizing gas delivery shall have the pressure sensors installed on the _
side of the individual room zone valve box assembly. 99-52 5.1.9.4.4 (1) - Answer-
Source or Patient
Computer systems used as a master alarm shall be continuously attended or shall
provide remote signaling of responsible parties through _. 99-52 5.1.9.3.1 (2) - Answer-
Pagers, telephones, auto-dials - any of the above.
High Carbon monoxide level in the medical air system shall indicate at each local alarm
site. At what level shall the alarm activate? 99-53 5.1.9.5.4 - Answer-10 ppm or higher
Ina cylinder supplied medical gas system with reserve, which alarm system(s) will
activate when, or just before, the changeover to the reserve supply occurs? - Answer-
Local and master
Category 1 spaces other than anesthetizing locations shall be located on which side of
the zone valve box assemblies? 99-52 5.1.9.4.4 (1) - Answer-Patient or use side
Initiating devices is remote from the building and the wiring is to run underground to a
junction box. How many sets of wires shall be permitted? 99-51 5.1.9.2.3.1 - Answer-
Single Set
One of the master alarm panels shall be located in the office or work space of the
individual responsible for the maintenance of the medical gas piping system. Where
Voordelen van het kopen van samenvattingen bij Stuvia op een rij:
Verzekerd van kwaliteit door reviews
Stuvia-klanten hebben meer dan 700.000 samenvattingen beoordeeld. Zo weet je zeker dat je de beste documenten koopt!
Snel en makkelijk kopen
Je betaalt supersnel en eenmalig met iDeal, creditcard of Stuvia-tegoed voor de samenvatting. Zonder lidmaatschap.
Focus op de essentie
Samenvattingen worden geschreven voor en door anderen. Daarom zijn de samenvattingen altijd betrouwbaar en actueel. Zo kom je snel tot de kern!
Veelgestelde vragen
Wat krijg ik als ik dit document koop?
Je krijgt een PDF, die direct beschikbaar is na je aankoop. Het gekochte document is altijd, overal en oneindig toegankelijk via je profiel.
Tevredenheidsgarantie: hoe werkt dat?
Onze tevredenheidsgarantie zorgt ervoor dat je altijd een studiedocument vindt dat goed bij je past. Je vult een formulier in en onze klantenservice regelt de rest.
Van wie koop ik deze samenvatting?
Stuvia is een marktplaats, je koop dit document dus niet van ons, maar van verkoper lectknancy. Stuvia faciliteert de betaling aan de verkoper.
Zit ik meteen vast aan een abonnement?
Nee, je koopt alleen deze samenvatting voor €10,28. Je zit daarna nergens aan vast.